ETH Logo PSI Logo

High-energy sources towards 2D applications

Goals and motivation of the project

  • feasibility of phase contrast on tubes with \(>100\) kVp
  • typical environment for medical, security and material science applications
  • characterize physical quantities

Outline

  • special thanks to Gordan, Thomas and Tine
  • technical studies
    • evaluation of KIT gratings
    • pilatus detector
  • applications
    • biomedical: cartilage
    • biomedical: lungs
    • materials: powders
    • materials: cement hydration

KIT gratings with Maria

  • various grating sets for 30 and 45 \(\pi\) shift
  • short summary: gratings are ok, homogeneity and absorption thickness are the main concerns.
  • \(\lt 100\) μm gold forces the smaller focal spot (1 mm)
  • visibility averaging 8-10% for most sets
  • best configuration reaching 15%

Best configuration

  • 100 μm G0, 140 μm G2, \(\pi\) shift 45 keV
  • eiger 1M detector
  • 80 kVp, 1 mm focal spot
Best visibility

Conclusions for our fabrication team

  • 80 μm G0, 140 μm G2, \(\pi\) shift 45 keV
  • pilatus CdTe detector
Visibility energy
  • increasing from 80 to 100 μm is very significant
  • this is also the very minimum absorption thickness for the setup